Resumen de Development and calibration of automatic real-time environmental radioactivity monitors using gamma-ray spectrometry

Ramon Casanovas Alegre

  • This doctoral thesis presents a collection of six scientific papers, which have been published in peer-reviewed scientific journals, in the field of real-time environmental radioactivity detection. After the contributions of this thesis, the automatic real-time environmental radiation surveillance network of Catalonia provides new and better radiological information. This was achieved thanks to the development and calibration of three types of radioactivity monitors using gamma-ray spectrometry, either with NaI(Tl) or LaBr3(Ce) detectors, which permit the real-time identification and quantification of radioactive isotope content in water and in air. In the thesis, details on general calibration methodologies can be found. The mentioned developments as well as their specific calibrations are also discussed. For each monitor, an evaluation of their measurement capabilities is also performed.