Comprehensive study of mechanical power in controlled mechanical ventilation: Prevalence of elevated mechanical power and component analysis

    1. [1] Hospital Universitario Marqués de Valdecilla

      Hospital Universitario Marqués de Valdecilla

      Santander, España

    2. [2] Hospital Universitario Central de Asturias

      Hospital Universitario Central de Asturias

      Oviedo, España

    3. [3] Hospital Universitario del Henares

      Hospital Universitario del Henares

      Coslada, España

    4. [4] Hospital Biprovincial Quillota-Petorca (Chile)
    5. [5] St. James's University Hospital (Dublin, Ireland)
    6. [6] Hélvio Auto Hospital (Alagoas, Brazil)
    7. [7] Hospital Universitario de Cabueñes (Gijón)
    8. [8] Hospital Universitario de Álava (Vitoria-Gasteiz)
    9. [9] Clinic University Hospital of Barcelona (Barcelona)
    10. [10] IRCCS Humanitas Research Hospital (Rozzano, Milano)
    11. [11] Hospital Universitari i Politecnic La Fe (Valencia)
  • Localización: Medicina intensiva, ISSN-e 1578-6749, ISSN 0210-5691, Vol. 48, Nº. 3, 2024, págs. 155-164
  • Idioma: inglés
  • Títulos paralelos:
    • Estudio Integral de la Potencia Mecánica en la Ventilación Mecánica Controlada: prevalencia de Potencia Mecánica Elevada y Análisis de Componentes
  • Enlaces
  • Resumen
    • English

      Objective To determine the prevalence of elevated mechanical power (MP) values (>17J/min) used in routine clinical practice.

      Design Observational, descriptive, cross-sectional, analytical, multicenter, international study conducted on November 21, 2019, from 8:00 AM to 3:00 PM. NCT03936231.

      Setting One hundred thirty-three Critical Care Units.

      Patients Patients receiving invasive mechanical ventilation for any cause.

      Interventions None.

      Main variables of interest Mechanical power.

      Results A population of 372 patients was analyzed. PM was significantly higher in patients under pressure-controlled ventilation (PC) compared to volume-controlled ventilation (VC) (19.20±8.44J/min vs. 16.01±6.88J/min; p<0.001), but the percentage of patients with PM>17J/min was not different (41% vs. 35%, respectively; p=0.382). The best models according to AICcw expressing PM for patients in VC are described as follows: Surrogate Strain (Driving Pressure) + PEEP+Surrogate Strain Rate (PEEP/Flow Ratio) + Respiratory Rate. For patients in PC, it is defined as: Surrogate Strain (Expiratory Tidal Volume/PEEP) + PEEP+Surrogate Strain Rate (Surrogate Strain/Ti) + Respiratory Rate+Expiratory Tidal Volume+Ti.

      Conclusions A substantial proportion of mechanically ventilated patients may be at risk of experiencing elevated levels of mechanical power. Despite observed differences in mechanical power values between VC and PC ventilation, they did not result in a significant disparity in the prevalence of high mechanical power values.

    • English

      Objetivo Determinar la prevalencia de valores elevados de potencia mecánica (PM) (>17J/min) utilizados en la práctica clínica habitual.

      Diseño estudio observacional, descriptivo de corte transversal, analítico, multicéntrico e internacional, realizado el 21 de noviembre de 2019 en horario de 8 a 15 horas. NCT03936231.

      Ámbito Ciento treinta y tres Unidad de Cuidados Críticos.

      Pacientes pacientes que recibirán ventilación mecánica por cualquier causa.

      Intervenciones ninguna Variables de interés principales Potencia mecánica.

      Resultados se analizaron 372 enfermos. La PM fue significativamente mayor en pacientes en ventilación controlada por presión (PC) que en ventilación controlada por volumen (VC) (19,20+8,44J/min frente a 16,01+6,88J/min; p<0,001), pero el porcentaje de pacientes con PM>17J/min no fue diferente (41% frente a 35% respectivamente; p=0,382). Los mejores modelos según AICcw que expresan la PM para los enfermos en VC se decribe como: Strain subrogante (Presión de conducción) + PEEP+Strain Rate subrogante (PEEP/cociente de flujo) + Frecuencia respiratoria. Para los enfermos en PC se define como: Strain subrogante (Volumen tidal expiratorio/PEEP) + PEEP+Strain Rate subrogante (Strain subrogante/Ti) + Frecuencia respiratoria+Expiratory Tidal Volumen+Ti.

      Conclusiones Gran parte de los pacientes en ventilación mecánica en condiciones de práctica clínica habitual reciben niveles de potencia mecánica peligrosos. A pesar de las diferencias observadas en los valores de potencia mecánica entre la ventilación VC y PC, este porcentaje de riesgo fue similar en PC y VC.