High performance of the automated ADVIA Centaur Systems SARS-CoV-2 Antigen Assay in nasopharyngeal samples with high viral load

    1. [1] Hospital Clínico San Carlos de Madrid

      Hospital Clínico San Carlos de Madrid

      Madrid, España

    2. [2] Universidad Complutense de Madrid

      Universidad Complutense de Madrid

      Madrid, España

    3. [3] Universidad Complutense Madrid
  • Localización: International microbiology: official journal of the Spanish Society for Microbiology, ISSN 1139-6709, Vol. 26, Nº. 3, 2023, págs. 471-474
  • Idioma: inglés
  • Enlaces
  • Resumen
    • ADVIA Centaur SARS-CoV-2 Antigen (COV2Ag) Assay (Siemens Healthineers) was evaluated for SARS-CoV-2 detection. A total of 141 nasopharyngeal samples were analyzed by this technique and results were compared with those obtained by quantitative reverse-transcription polymerase chain reaction (RT-PCR). The overall sensitivity and specificity of the test were 68.70% and 70%, respectively. Regarding cycle threshold (Ct) values, the COV2Ag test showed a sensitivity of 93.75% and 100% for nasopharyngeal samples with Ct < 25 and < 20, respectively. ADVIA Centaur COV2Ag Assay is a useful, automated, and rapid technique for early SARS-CoV-2 diagnosis and isolation of the infected individuals, avoiding its transmission.