Ayuda
Ir al contenido

Dialnet


Lens-vault analysis and its correlation with other biometric parameters using swept-source OCT

    1. [1] Universitat de València

      Universitat de València

      Valencia, España

    2. [2] Oftalvist Clinic, Alicante, Spain
  • Localización: Journal of Optometry: peer-reviewed Journal of the Spanish General Council of Optometry, ISSN-e 1888-4296, Vol. 15, Nº. 1, 2022, págs. 88-99
  • Idioma: inglés
  • Enlaces
  • Resumen
    • Purpose To measure lens vault (LV) and to assess its correlation with various ocular parameters in healthy eyes, using for all measurements the same high-resolution swept-source optical coherence tomographer (SS-OCT).

      Methods We prospectively recruited 67 Caucasian healthy patients whose mean age was 41.9 ± 12.4 years; only their right eye was included in the study. Data were all recorded with the ANTERION SS-OCT and comprised, for each patient, 5 consecutive measurements of LV, anterior chamber depth (ACD), lens thickness (LT), axial length (AL), white-to-white (WTW) distance, central corneal thickness (CCT), anterior chamber volume (ACV) and spur-to-spur (STS) distance.

      Results Mean LV was 0.26 ± 0.23 mm (ranging from -0.24 to 0.78 mm). Data analysis revealed a statistically significant negative correlation between LV and ACD (R=-0.80, p < 0.001), AL (R = -0.36, p = 0.002), and ACV (R = -0.68, p < 0.001), and a positive correlation between LV and LT (R = 0.67, p < 0.001), and age (R = 0.53, p < 0.001). In contrast, no statistically significant correlation was found between LV and WTW (R=-0.17, p = 0.15), CCT (R = 0.11, p = 0.36) or STS (R=-0.10, p = 0.41).

      Conclusions Taking into account our findings about intra-parameter correlation levels, we believe that LV should be measured and analyzed together with other ocular parameters in clinical routine practice both for diagnosis and for some refractive surgeries.


Fundación Dialnet

Dialnet Plus

  • Más información sobre Dialnet Plus

Opciones de compartir

Opciones de entorno