Modelo matemático de letalidad sobre bacillus arthophaeus en procesos de esterilización de dispositivos médicos con óxido de etileno

  • Autores: Natalia Jaramillq-Ruiz, Pedronel Araque Marín, Yesid Mqntqya-Góez
  • Localización: Revista EIA, ISSN-e 1794-1237, Vol. 14, Nº. 28, 2017, págs. 45-55
  • Idioma: español
  • Títulos paralelos:
    • Modelo matemático de letalidade contra bacillus atrophaeus em processos de esterilização de dispositivos médicos com óxido de etileno
    • Mathematical model of lethality against bacillus atrophaeus in sterilization processes of medical devices with ethylene oxide
  • Enlaces
  • Resumen
    • español

      RESUMEN La esterilización de dispositivos médicos con óxido de etileno tiene como objetivo eliminar toda forma de vida que pueda tener contacto con los tejidos corporales, destruyendo la contaminación microbiana que podría dar lugar a la transmisión de enfermedades. La esterilización con óxido de etileno se realiza a bajas temperaturas y se aplica a los dispositivos médicos que no pueden ser esterilizados con vapor de agua. A demás de ser un proceso costoso, la esterilización con óxido de etileno presenta algunas desventajas al ser implementada con ciclos de tiempos prolongados, convirtiéndose en un factor de riesgo para los pacientes y el personal médico. El presente estudio propone un modelo matemático para medir y predecir la tasa de letalidad en procesos de esterilización de dispositivos médicos con OE en función de la concentración de óxido de etileno, humedad relativa, presión, temperatura de trabajo y el tiempo de exposición del dispositivo con óxido de etileno. El modelo matemático propuesto describe un comportamiento cinético de primer orden, para un numero de esporas de Bacillus arthophaeus de 3.5x106, una concentración de óxido de etileno de 700 ppm, una temperatura ambiente de 23 °C, una temperatura de trabajo de 60 °C, una humedad relativa del 60 % y una presión de trabajo 0.490 atm.

    • English

      ABSTRACT Sterilization of medical devices with ethylene oxide aims to eliminate all forms of life that may have contact with body tissues, destroying microbial contamination that could lead to disease transmission. Ethylene oxide sterilization is performed at low temperatures and it's applied to medical devices that cannot be steam sterilized. Besides being an expensive process, Ethylene oxide sterilization has some disadvantages when implemented with prolonged time cycles, becoming a risk factor for patients and medical staff. This study proposes a mathematical model to measure and predict the lethal rate in the process of ethylene oxide sterilization of medical devices based on its concentration, relative humidity, pressure, working temperature and exposure time of the device to the ethylene oxide. The proposed mathematical model describes a first-order kinetic behavior, for a number of spores of Bacillus arthophaeus of 3.5x106, a concentration of ethylene oxide of 700 ppm, an ambient temperature of 23°C, a working temperature of 60 °C, a relative humidity of 60 % and a pressure 0.490atm.

    • português

      RESUMO Esterilização de dispositivos médicos com óxido de etileno visa eliminar todas as formas de vida que possa ter contacto com os tecidos do corpo, destruindo a contaminação microbiana que poderia levar a transmissão da doença. esterilização por óxido de etileno é realizada a baixas temperaturas e é aplicado a dispositivos médicos que não podem ser esterilizados por vapor. Além de ser um processo caro, esterilização por óxido de etileno tem algumas desvantagens quando implementado com ciclos de tempo prolongados, tornando-se um fator de risco para os pacientes e pessoal médico. Este estudo propõe um modelo matemático para medir e prever a taxa letal no processo de esterilização por óxido de etileno de dispositivos médicos com base na sua concentração, a humidade relativa, a pressão, a temperatura de trabalho e tempo de exposição do dispositivo ao óxido de etileno. O modelo matemático proposto descreve um comportamento cinético de primeira ordem, para um número de esporos de Bacillus arthophaeus de 3.5x106, uma concentração de óxido de etileno de 700 ppm, com uma temperatura ambiente de 23 ° C, uma temperatura de funcionamento de 60 ° C, uma humidade relativa de 60% e um 0.490atm pressão.