Resumen de Evaluation of concordance among three cardiac output measurement techniques in adult patients during cardiovascular surgery postoperative care

L. Muñoz, Ana Luisa Velandia Mora, L.E. Reyes, I. Arevalo-Rodríguez, C. Mejía, D. Asprilla, D.V. Uribe, J.J. Arevalo

  • español

    Introducción El cálculo del gasto cardiaco se realiza por termodilución, y su principal desventaja es el carácter invasivo. La ecocardiografía transesofágica (ETE) representa una alternativa dinámica y funcional a la termodilución.

    Objetivo Analizar la concordancia entre dos métodos de ETE y termodilución para la evaluación del gasto cardiaco.

    Métodos Estudio observacional de concordancia en pacientes de cirugía cardiovascular con requerimiento de catéter de arteria pulmonar. Se realizó medición de gasto cardiaco por ETE en anillo mitral (AM) y en el tracto de salida del ventrículo izquierdo (TSVI). Los resultados se compararon con la termodilución. La concordancia fue evaluada por el coeficiente de correlación concordancia de Lin y analizada por el método de Bland-Altman. Los análisis estadísticos se realizaron en STATA 13.0.

    Resultados Se incluyeron 25 pacientes. El 52% fueron hombres, con mediana de edad de 63años y fracción de eyección del 35%. La mediana de gasto cardiaco por termodilución, AM y TSVI fue de 3,25, de 3,46 y de 8,4L/min, respectivamente. Se encontraron diferentes valores entre termodilución y AM (concordancia de Lin=0,071; IC95%: −0,009 a 0,151), así como entre termodilución y TSVI (concordancia de Lin=0,232; IC95%: −0,12 a 0,537). El análisis de Bland-Altman muestra una diferencia entre la medición por AM y termodilución importante (DM=−0,408; Bland-Altman Limits=−0,809 a −0,007), así como entre las dos medidas por ETE (DM=0,007; Bland-Altman Limits=−0,441 a 0,428).

    Conclusión Los resultados en la medición del gasto cardiaco por doppler y ETE bidimensional tanto a nivel del anillo mitral como del TSVI no son concordantes con la termodilución.

  • English

    Introduction The standard method for cardiac output measuring is thermodilution although it is an invasive technique. Transesophageal Echocardiography (TEE) offers a dynamic and functional alternative to thermodilution.

    Objective Analyze concordance between two TEE methods and thermodilution for cardiac output assessment.

    Methods Observational concordance study in cardiovascular surgery patients that required pulmonary artery catheter. TEE cardiac output measurement at both mitral annulus (MA) and left ventricle outflow tract (LVOT) were performed. Results were compared with thermodilution. Correlation was evaluated by Lin's concordance correlation coefficient and Bland–Altman analysis. Statistical analysis was undertaken in STATA 13.0.

    Results Twenty-five patients were enrolled. Fifty two percent of patients were male, median age and ejection fraction was 63 years and 35% respectively. Median thermodilution, LVOT and MA -measured cardiac output was 3.25 L/min, 3.46 L/min and 8.4 L/min respectively. Different values between thermodilution and MA measurements were found (Lin concordance=0.071; Confidence Interval 95%=−0.009 to 0.151; Spearman's correlation=0.22) as values between thermodilution and LVOT (Lin concordance=0.232; Confidence Interval 95%=−0.12 a 0.537; Spearman's correlation 0.28). Bland–Altman analysis showed greater difference between MA measurements and thermodilution (DM=−0.408; Bland–Altman Limits=−0.809 to −0.007), than the other echocardiographic findings (DM=0.007; Bland–Altman Limits=−0.441 to 0.428).

    Conclusion Results from cardiac output measurement by doppler and 2D-TEE on both MA and LVOT do not correlate with those obtained by thermodilution.