Ayuda
Ir al contenido

Dialnet


Why We Need Postmortem Analysis of Cardiac Implantable Electronic Devices

    1. [1] University of Zurich

      University of Zurich

      Zürich, Suiza

    2. [2] University Hospital Zurich Division of Trauma Surgery, Department of Surgery Zurich Switzerland
  • Localización: Journal of forensic sciences, ISSN-e 1556-4029, ISSN 0022-1198, Vol. 61, Nº. 4, 2016, págs. 988-992
  • Idioma: inglés
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • The prevalence of cardiac implantable electronic devices (CIEDs), pacemakers and implantable cardioverter defibrillators (ICDs) is increasing. However, postmortem analysis of CIEDs is not performed routinely. Fourteen consecutive CIEDs were analyzed. The indication for and date of implantation, technical data, CIED reprogramming, heart rhythm disturbances, patient demographics and medical consultations were investigated. Death during the first year after implantation was seen in 54%, whereof 71% consulted a physician within 10 days before death. The time of death was attributed to a particular day in 29%. There was a relationship between CIEDs and cause/manner of death in 50%. Although limited by a small sample size, this study advocates the routine postmortem CIED analysis for forensic and clinical purposes in selected cases. Patients with CIEDs seem to show an increased risk of death during the first year after implantation. The analysis of CIEDs can be helpful in evaluating the time/cause/manner of death.