Comparison of blood pressure measurements of anesthetized dogs obtained noninvasively with a cylindrical blood pressure cuff and an anatomically modified conical blood pressure cuff Sara J. Ramos, DVM; Anderson F. da Cunha, DVM, MS; Michelle Domingues, DVM; Amanda M. Shelby, BS; Rhett W. Stout, DVM, PhD; Mark J. Acierno, MBA, DVM Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803. (Ramos, da Cunha, Domingues, Shelby, Acierno); Division of Laboratory Animal Medicine, School of Veterinary Medicine, Louisiana State University, Baton Rouge, LA 70803. (Stout) Drs. Ramos and Domingues were third-year veterinary students at the time of the study.
Address correspondence to Dr. da Cunha (adacun1@lsu.edu).
OBJECTIVE To compare blood pressure measured noninvasively with an oscillometric device that involved use of a novel conical cuff and a traditional cylindrical blood pressure cuff.
ANIMALS 17 adult hound-type dogs.
PROCEDURES Dogs were anesthetized, and a 20-gauge, 1.5-inch catheter was inserted in the median sacral artery. The catheter was attached to a pressure transducer via fluid-filled noncompliant tubing, and direct blood pressure was recorded with a multifunction monitor. A specially fabricated conical cuff was placed on the antebrachium. Four sets of direct and indirect blood pressure measurements were simultaneously collected every 2 minutes. Four sets of measurements were then obtained by use of a cylindrical cuff.
RESULTS The cylindrical cuff met American College of Veterinary Internal Medicine consensus guidelines for validation of indirect blood pressure measurements for mean arterial blood pressure (MAP), systolic arterial blood pressure (SAP), and diastolic arterial blood pressure (DAP). The conical cuff met the consensus guidelines for difference of paired measurements, SD, and percentages of measurements within 10 and 20 mm Hg of the value for the reference method, but it failed a correlation analysis. In addition, although bias for the conical cuff was less than that for the cylindrical cuff for SAP, MAP, and DAP measurements, the limits of agreement for the conical cuff were wider than those for the cylindrical cuff for SAP and MAP measurements.
CONCLUSIONS AND CLINICAL RELEVANCE On the basis of results of this study, use of a conical cuff for oscillometric blood pressure measurement cannot be recommended.