Ayuda
Ir al contenido

Dialnet


Resumen de Influence of the Implant Drill Design and Sequence on Temperature Changes During Site Preparation

Alberto Barlattani, Gianpaolo Sannino, Paolo Capparè, Enrico F. Gherlone

  • Purpose: The purpose of this study was to compare bone temperature changes during implant drilling with two drill designs employed in three different drilling sequences. Materials and Methods: Two implant drill designs and three drilling sequences were evaluated in vitro using artificial bone cylinders. The evaluated drills were different only in the cutting-surface length (control, 16 mm; test, 4 mm). Three drilling sequences (control A, test B1, and test B2) were evaluated with and without irrigation. Temperatures were measured with thermocouple technology. The temperature changes generated by the final drill of each sequence were recorded as the experimental results and were subjected to the Student t test. Results: There were statistically significant differences in temperature changes when comparing the control group A with the test groups B1 (P = .001) and B2 (P = .01) during drilling without coolant. The mean temperature changes were 12.4°C, 6.5°C, and 13.7°C for groups A, B1, and B2, respectively. The Student t test showed statistically significant differences between temperature changes of the control group A and the test groups B1 (P < .01) and B2 (P < .05) during drilling with coolant. The mean temperature changes were 0.9°C, 0.7°C, and 1.9°C for groups A, B1, and B2, respectively. Conclusion: Reduction in length of the cutting surface of the drill may limit frictional heat. Drills with the same length of cutting surface may induce lower bone temperature changes, when considering a preliminary drilling step with a pilot drill.


Fundación Dialnet

Dialnet Plus

  • Más información sobre Dialnet Plus