Reproducibility of Facial Soft Tissue Thickness Measurements Using Cone-Beam CT Images According to the Measurement Methods

    1. [1] Chonnam National University

      Chonnam National University

      Corea del Sur

    2. [2] Seoul National University

      Seoul National University

      Corea del Sur

    3. [3] University of Dundee

      University of Dundee

      Reino Unido

  • Localización: Journal of forensic sciences, ISSN-e 1556-4029, ISSN 0022-1198, Vol. 60, Nº. 4, 2015, págs. 957-965
  • Idioma: inglés
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • The purpose of this study was to establish the reproducibility of facial soft tissue (ST) thickness measurements by comparing threedifferent measurement methods applied at 32 landmarks on three-dimensional cone-beam computed tomography (CBCT) images. Two observers carried out the measurements of facial ST thickness of 20 adult subjects using CBCT scan data, and inter- and intra-observer reproducibilities were evaluated. The measurement method of “perpendicular to bone” resulted in high inter- and intra-observer reproducibility at all 32 landmarks. In contrast, the “perpendicular to skin” method and “direct” method, which measures a distance between one point on bone and the other point on skin, presented low reproducibility. The results indicate that reproducibility could be increased by identifying the landmarks on hard tissue images, rather than onST images, and the landmark description used in this study can be used in the establishment of reliable tissue depth data using CBCT images.