Ayuda
Ir al contenido

Dialnet


Quality control of the breast ca treatments on HDR breachytherapy with TLD-100

  • Autores: F. Torres Hoyos, Nicolás de la Espriella Vélez, A. Sánchez Caraballo
  • Localización: Revista Mexicana de Física, ISSN-e 0035-001X, Vol. 60, Nº. 6, 2014, págs. 400-413
  • Idioma: inglés
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • An anthropomorphic Phantom, a female trunk, was built with a natural bone structure and experimental material conted, glycerin and waterbased material called JJT to build soft tissue equivalent to the muscle of human tissue, and a polymer (styrofoam) to build the long as critical organ to simulate the treatment of breast cancer, with high dose rate brachytherapy (HDR) and sources of Ir-192. The treatments were planned and calculated for the critical organ: Lung, and injury of 2 cm in diameter in breast with MicroSelectron HDK system and the software Plato Brachytherapy V 14.1 of the Nucletron (Netherlands) which uses the standard protocol of radiotherapy for brachytherapy treatments. The dose experimentally measured with dosimeters TLD-100 LIF: Mg: Ti, which were previously calibrated, were placed in the same positions obtained for calculated treatments, using the standard simulator, and the experimental with TLD-100, show a high concordance, as they are on average +/- 1.1% making process becomes in a quality control of this type of treatments.


Fundación Dialnet

Dialnet Plus

  • Más información sobre Dialnet Plus

Opciones de compartir

Opciones de entorno