Tips of stainless steel K-files from four different manufacturers (Kerr, Maillefer, Moyco and CCCord) were investigated microscopically to evaluate the dimensional conformity of endodontic instruments with respect to ANSI/ADA specification no.28 and 1SO 3630-1 specification. The tip length, tip angle and the diameter D, were measured using an optic microscope. The results of measurements of the tip dimensions revealed a significant number of instruments not in compliance with specifications, especially for the tip lengths, which were found values often lower than or close to the inferior prescribed limits. Most D, values were found within the tolerance limits for all the brands studied, although few files showed the exact nominal size.
Com o objetivo de avaliar a conformidade dimensional dos instrumentos endodônticos manuais em relação as especificações número 28 da ANSI/ADA e ISO 3630-1, foi realizada uma investigação microscópica na ponta de limas manuais de ago inoxidável tipo K de quatro diferentes marcas comerciais (Kerr, Maillefer, Moyco e CCCord). Foram medidos num microscópio ótico de medição: o diâmetro D, o ângulo e o comprimento da ponta. Os resultados das medidas de dimensões das pontas revelaram um número significativo de limas fora das especificações, principalmente para o comprimento da ponta. Os valores de D, apesar de se encontrarem em sua maioria, para todas as marcas pesquisadas, dentro dos limites de tolerância, raramente exibiram o valor nominal exato.