Ayuda
Ir al contenido

Dialnet


Resumen de Copperii-selective electrodes based on 1-furoyl-3,3-diethylthiourea as a neutral carrier

M. de los A Arada Pérez, Santiago Landazury Yanes, M. Cardona, L Aguilera, M. Yazdani Pedram

  • 1-furoyl-3,3-diethylthiourea was used as a novel neutral carrier in ion selective electrode for copperII detection and determination. The Cu2+ selective plasticized polyvinyl chloride (PVC) membrane electrode exhibited reversible potential responses with Nernstian slope of 29.34± 0.6 mV/decade for Cu2+ ions in the concentration range from 1.0x10-7 to 1.0x10-2 mol/dm³ with a practical detection limit (PDL) of 3.84x10-7 and lower limit of linear response (LLLR) of 6.15x10-7 mol/dm³ and alife-time of 20 days. The response time, pH effect and other characteristics of the electrodes were also studied.