Concordancia del Tonómetro de Goldmann y Corvis ST en pacientes afectos de glaucoma intervenidos de cirugía refractiva LASIK

    1. [1] Hospital Universitari Germans Trias i Pujol

      Hospital Universitari Germans Trias i Pujol

      Barcelona, España

  • Localización: Annals d'oftalmologia: òrgan de les Societats d'Oftalmologia de Catalunya, Valencia i Balears, ISSN-e 1133-7737, Vol. 29, Nº. 1, 2021, págs. 3-6
  • Idioma: español
  • Títulos paralelos:
    • Goldman tonometer and Corvis ST concordancy, in patients with glaucoma who underwent LASIK refractive surgery
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • español

      Introducción: Medir la presión intraocular de forma fiable en pacientes intervenidos de cirugía refractiva corneal es un desafío por la alteración de la morfología y biomecánica corneal.

      Material y métodos: Estudio observacional que compara la correlación entre presiones intraoculares determinadas por tonometría de Corvis ST sin corrección, corregida biomecánicamente (bIOP) y tonometría de Goldmann (TAG) sin corrección y corregida según paquimetría. Se estudian un total de 12 ojos intervenidos de LASIK. Se utiliza el coeficiente de correlación intraclase (Bland-Altman) para estimar la correlación de la presión intraocular estimada por ambos métodos y t-student para saber si las diferencias han sido estadísticamente significativas (p < 0,05).

      Resultados: La correlación es positiva en todos los casos. Existe una mayor correlación entre la TAG y la TAG corregida según paquimetría, llegando a un coeficiente de 0,7. En cuanto a las diferencias de media en mmHg se aprecian diferencias estadísticamente significativas cuando comparamos la TAG corregida según paquimetría y los otros tres grupos antes descritos.

      Conclusión: En pacientes glaucomatosos intervenidos de cirugía refractiva corneal se debe tener precaución en la medición de la PIO a través de la TAG, y apoyarnos en el uso de métodos complementarios.

    • English

      Introduction: Measuring the intraocular pressure in patients who underwent corneal refractive surgery in a reliably form is challenging, due to the alteration of the corneal morphology and biomechanics.

      Material and methods: Observational study that compares the correlation between the Corvis ST tonometry without correction, biomechanically corrected and Goldmann tonometry (GAT) without correction and corrected by pachymetry. A total of 12 eyes that underwent LASIK surgery were studied. Bland- Altman correlation coefficient was used to estimate the two methods correlation and t-student to determinate if exist statistic differences in the average of measures.

      Results: It was found a positive correlation in all the methods. Between the GAT and the GAT with pachymetry correction it was found the more positive correlation arriving at 0.7. As the comparison between measures averages in mmHg, significant statistic differences was found between corrected GAT by pachymetry and the other three groups.

      Conclusion: In glaucoma patients with refractive corneal surgery, it must be careful not to rely entirely on the measurement of IOP through GAT and help us with complementary methods.

    • català

      Introducció: Mesurar la pressió intraocular de manera fiable en pacients intervinguts de cirurgia refractiva cornial es un repte per l’alteració de la morfologia i biomecànica de la còrnia.

      Material i mètodes: Estudi observacional, que compara la correlació entre pressions intraoculars determinades per tonometria de Corvis ST sense correcció, corregida biomecànicamene (bIOP), i tonometria de Goldmann (TAG) sense correcció i corregida segons paquimetria. S’estudien un total de 12 ulls intervinguts de LASIK. S’utilitza el coeficient de correlació intraclase (Bland-Altman) per a estimar la correlació de la pressió intraocular estimada per ambdós mètodes i la t-student per a saber si les diferencies han estat estadísticament significatives (p>0,05).

      Resultats: La correlació es positiva en tots els casos. Existeix una major correlació entre la TAG sense correcció i la pressió mesurada amb la TAG corregida segons paquimetria, arribant a un coeficient de 0,7. En quant a les diferencies de mesura en mmHg s’aprecien diferencies estadísticament significatives quan comparem la TAG corregida segons paquimetria i els altres tres grups anteriorment descrits.

      Conclusió: En pacients glaucomatosos intervinguts de cirurgia refractiva corneal s’ha de tenir precaució amb la mesura de la PIO a través de la TAG i, recolzar-nos en l’ús de mètodes complementaris.