Planning target volume assessment in lung tumors during 3D conformal radiotherapy by means of an aSi electronic portal imaging device in cine mode

  • Autores: R. Caivano, S. Clemente, A. Fiorentino, C. Chiumento, P. Pedicini, G. Califano, M. Cozzolino, V. Fusco
  • Localización: Clinical & translational oncology, ISSN 1699-048X, Vol. 15, Nº. 8, 2013, págs. 638-642
  • Idioma: inglés
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • Purpose The major uncertainties in treating lung cancer are the repositioning errors and respiratory lung tumor motion. Typically, margins are added to the clinical target volume (CTV) to obtain a planning target volume (PTV) allowing the accommodation of such uncertainties. We want to test a new technique to assess the adequacy of the chosen PTV using an aSi electronic portal imaging device (EPID).

      Methods Four patients affected by lung cancer and treated by radical 3D conformal radiotherapy (3DRT) were studied. During treatment the EPID was used in cine mode acquisition: acquired images were used to the aim.

      Results and conclusions Treatment monitoring with an EPID in cine mode is shown to be a clinically feasible and useful tool.